IDN-02 Neonate intensive care incubator
The Infant intensive care incubator IDN-02 is intended for nursing and carrying out of effective resuscitation of premature and weakened babies in departments of neonates pathology, resuscitation sections, intensive therapy wards of specialized medical institutions.
The IDN-02 incubator consists of infant module with double walls and transport trolley for free movement. IDN-02 provides warming environment, air humidification and oxygen supply inside the infant module.
Technical characteristics
Adjustment of temperature inside infant module. from 25°C to 39°C
Adjustment of temperature over neonate’s skin from 34°C to 39°C
Measurement step 0,1
Admissible error 0,3
Relative humidity adjustment. from 60 to 85%
Oxygen concentration adjustment from 21 to 95%
Vertical displacement (lift) of infant module 200 mm
Infant module tilt angle 9°±1°
Continuous operation time more than 72 hours
Level of sound in infant module 50 dBA
Concentration of CO, in module not more than 0,4%
Class of safety llb