PG NEST incubator’s air circuit supplies quiet and comfortable ambient, helping to reduce external noise. The inner temperature is monitored and controlled on various points to ensure accuracy and safety, it supports 39 different alarms and 16 error messages. Conditions of air, humidity, oxygen sensor, air flow probes and sensor box fan are monitored by an accurate alarms system. A proper monitoring system alerts the operator about the working conditions of technical and physiological parameters. A back up battery supports the working conditions for at least 30 minutes.

 

TECHNICAL SPECIFICATIONS

PHYSICAL ATTRIBUTES

Height: 126.1cm to 146.1cm

Width: 99.7cm Depth: 57.7cm Weigh: 80kg Package Dimensions: 107*68*155 cm Tubing access ports: 10 Access door size: 18cm x 13cm Mattress to hood height: 48.0mm Soft bed mattress size: 73.6cm x38.6cm x1.8cm

DISPLAY

8” LCD display, interactive operation interface

TEMPERATURE CONTROL MODES

Temperature control modes: Skin and air temperature control mode Air mode control temperature range: 20° C-37° C Air mode control override temperature range: 37° C-39° C Air mode control accuracy: ≤±1.0°C Skin mode control temperature range: 34° C-37° C Skin mode control override temperature range: 37° C-38° C Skin temperature accuracy: ±0.3°C Dual-skin temperature monitoring: Yes

PERFORMANCE

Air flow velocity across mattress: < 10 cm/sec Temperature rise time at 22 °C (72 °F) ambient: <20min (From 22 °C, 50% RH, to 35 °C) Temperature variability: < 0.5 °C Temperature overshoot: < 0.5 °C maximum Temperature uniformity with a level mattress: < 0.8 °C Operating noise level in hood: ≤45dBa Carbon Dioxide (CO2) Level: <0.5%

SERVO HUMIDITY OPTION

Humidity control range: 30%-95% RH Humidity control operating time without refilling: 24 hours Humidity control reservoir capacity: 1500ml Humidity display accuracy: ±5%RH Humidity control accuracy: ±5% RH

SERVO OXYGEN OPTION

Oxygen control range: 21 % to 65% Oxygen display accuracy of full scale: ± 2% Oxygen control accuracy (100% calibration): ± 3% Oxygen control accuracy (21% calibration): ± 5% Oxygen display resolution: 1%

SCALE OPTION

Weight range: 300 g to 8 kg Weight display resolution: 1 g Weight accuracy: ±10 g Operating Environment: 20° to 30° C Temperature: 10 to 95% RH Air Velocity: Up to 0.3 m/sec

Big storage cabinet and two sides of drawers. 1,5 detachable water reservoir keeps running up to 36 hours. Ergonomic handle. Three auxiliary power output. 4 castors with locking system.